Skip to main content Skip to search Skip to main navigation

Portavo 907 Multichannel MS Portable Multiparameter Transmitter | Memosens Sensors | Multichannel Function | Data Logger | Add-On Functions

Description

Portavo 907 Multichannel MS Portable Multiparameter Transmitter | Memosens Sensors | Multichannel Function | Data Logger | Add-On Functions

Portable multiparameter analyzer with multi-channel function for simultaneous measurement with Memosens sensors

  • Multiparameter:
    • pH
    • ORP
    • Contacting conductivity
    • Toroidal conductivity
    • Amperometric oxygen
    • Optical oxygen
    • Temperature
  • Oxygen measurement in liquids or in the gaseous phase
  • Multichannel function
  • GLP compliant
  • User management
  • New add-on functions, such as a new pH calibration procedure, user management, sensor verification, and calibration of the temperature detector, are available as options.
  • Digital Memosens sensors
  • Concentration measurement with toroidal conductivity sensors
  • Sturdy, practical, convenient
  • Li-ion rechargeable battery - USB chargeable

Portavo 907 Multichannel MS is the first multi-channel device in the Portavo product line. Any 2 Memosens sensors can be connected at the same time. In addition to a Memosens sensor, it is also possible to use the SE 340 digital optical oxygen sensor.

Many new features distinguish the Portavo 907 Multichannel MS for use in the pharmaceutical and biotech field. These include 

  • a new pH calibration procedure with a fixed process cycle
  • multi-level user management with access control
  • direct assignment of Memosens sensors to the device,which increases safety in operation.

Custom pH Calibration
Cal SOP

The new Cal SOP calibration procedure allows pH sensors to be checked with up to 3 calibration points. A further buffer is used as the verification buffer. The buffer set for each calibration point can be separately selected, thus also allowing their order to be determined.

Custom buffer solutions can be used, or choose from a list of commercially available buffer sets, e.g., CaliMat, NIST, and DIN. A maximum permissible deviation (Delta pH) is entered for the verification buffer.

Full Measured Value Recording
Data logger functionality for both sensors in use. Measurement data is saved and displayed directly in the device.

Security Package Included
User management

The Portavo 907 Multichannel MS’s professional user management regulates access to the device and the sensor.

  • Increased security for configuration, calibration, and measurement data
  • No unauthorized interventions during the operating cycle
  • Up to 4 user profiles can be set
  • Different access rights can be established

Depending on the user’s experience, the role profile can optionally be defined for configuration of the device and sensor or for sensor calibration. This clearly minimizes the risk of inadvertently changing settings.

Greater Reliability During Operation
Memosens sensors can be assigned directly to Portavo 907 Multichannel MS using the data stored in the sensor, such as

  • Sensor type
  • TAG
  • Group

Unambiguous assignment of the sensor to the device reduces the potential for errors. This ensures that only the right sensors are used for the selected measuring point.

We look forward to hearing from you!
Contact Us

Technical details

Additional functions: Sensor verification, Temperature Adjustment, User management
Channels: 2 channels
Data Logger: data logger with 10,000 entries
Display: QVGA TFT display with white backlighting
Measurement Parameter: Multiparameter
Sensor Connection: Memosens
permanently installed / portable: portable
Explosion protection: non Ex
Enclosure: PA12 GF30 + TPE
Power Supply: 1x Li-ion Rechargeable battery (via USB), 4x AA Mignon Alkaline batteries, 4x NiMH Rechargeable batteries

Downloads

Manuals

[14442] Quickstart Guide
Portavo 907 Multi
Version 04
[12673] User Manual
Software Paraly SW 112
Version 05
[13862] Safety Guide
Portavo 900 Series
Version 04
[16861] ChangeLog Firmware History
Portavo-907-908
[1968] User Manual
Portavo 907 Multi
Version 05

Brochures

[16192] Brochure
Portavo (Overview)

Data Sheets

Data sheet
Portavo 907 Multichannel MS
Product Description | Specifications | Product Range
[2022] Datasheet
Overview of Portavo

Certificates

[11539] EU Declaration of Conformity (DE, EN, FR)
Serie / Series / Série Portavo 900
No. EU211124A | 2021-11-24

Software

[12703] Software
Paraly SW 112 PC Software

Our service

Used in Process or Applications

Knick offers innovative process analyzers, high quality holders and fittings, and unique cleaning and calibration systems

Related Products

Relevant Blog Posts

Sturdy 2-Electrodes Conductivity Sensor SE625 with Coaxial Electrodes made of Titanium

The SE625 ensures reliable and accurate determination of the conductivity of ultrapure water in the semiconductor industry.

Combined pH- and ORP-Sensor SE555X/*-AMSN

Combined pH and ORP measurement in one sensor SE555 saves place and equipment costs for demanding requirements in the chemical industry, the food sector and for hygienic applications.

SE655X Digital and SE656X Digital - Digital Conductivity Sensors with Ex Certification

The SE655X digital and SE656X digital conductivity sensors with Memosens protocol are now also available with Ex certification.

New SE656N digital conductivity sensor

New toroidal Digital Conductivity Sensor with a large measuring range and consisting of PFA for using in aggressive media

How to get the sensor into the process

Sensors as the direct interface to the process must be exactly matched to the chemistry of the measured medium and to the prevailing pressure and temperature conditions.